Zou, Li, Liu, and Hu: Challenges in PT/INR harmonization across the total testing process

Introduction

The most well-established preanalytical determinants of prothrombin time/international normalized ratio (PT/INR) reliability originate from specimen collection. Proper venipuncture technique is essential to avoid local activation or contamination. Prolonged tourniquet application, traumatic venipuncture, incorrect collection order, and sampling through indwelling lines may compromise sample quality and introduce variability in clot-based assays (1, 2).

The International Normalized Ratio, introduced by the World Health Organization (WHO) in 1983, uses the International Sensitivity Index (ISI) to reduce thromboplastin-related differences and establish metrological traceability (3, 4). Despite being one of the most influential standardization tools in laboratory medicine, “standardization” has not fully translated into “harmonization” - clinically equivalent results across all methods and settings (5-8).

Analytical variability also stems from how clot formation is detected. Prothrombin time systems rely on mechanical or viscosity detection, or optical monitoring of light transmission changes during fibrin formation. These methods differ not only in physical basis but also in sensitivity to hemolysis, lipemia, icterus, hypofibrinogenemia, and other matrix abnormalities (9, 10). Even when the same specimen is analyzed with closely related reagent systems, inconsistencies may persist because the clotting endpoint is operationally assigned rather than universally observed. In optical systems, the reported PT may depend on signal thresholds, clotting-curve processing rules, and platform-specific endpoint algorithms (11-13). Such differences may have clinical consequences when PT/INR results are used for vitamin K antagonist dose adjustment, interpretation of direct oral anticoagulant exposure, liver-disease severity assessment, or comparison of results across institutions.

The aim of this narrative review is therefore to critically examine residual PT/INR non-equivalence across the total testing process. We focus on how preanalytical conditions, reagent-instrument interactions, endpoint-detection principles, postanalytical INR parameters, external quality assessment materials, and clinical context influence the degree to which PT/INR results can be considered clinically interchangeable. By linking these sources of variability to patient-management consequences, we seek to clarify why PT/INR harmonization requires more than correct application of the INR formula and why system-specific verification and context-aware interpretation remain necessary.

Prothrombin time and International Normalized Ratio

The expanding use of direct oral anticoagulants (DOACs) illustrates an important boundary of PT/INR harmonization beyond the vitamin K antagonist (VKA) model. Unlike VKAs, which reduce multiple vitamin K-dependent factors in a relatively coordinated manner, DOACs produce target-specific effects that reflect drug class, plasma concentration, and reagent-dependent analytical responsiveness rather than biologically standardized patterns (14-16). Accordingly, PT/INR results in DOAC-exposed samples should not be interpreted as standardized measures of anticoagulant intensity. The clinical magnitude of this divergence is substantial. Tripodi and colleagues demonstrated that plasma from patients with advanced liver disease produced INR differences of 15-20% when tested with reagents at opposite ends of the ISI spectrum (ISI 0.9 vs. ISI 1.7), despite equivalent calibration traceability (17).

International Normalized Ratio was introduced to reduce PT variability by transforming it into a standardized expression: INR = (PTpatient / MNPT)ISI, where MNPT denotes the mean normal prothrombin time and ISI denotes the International Sensitivity Index (3, 4). This transformation improves comparability mainly in patients receiving VKAs, in whom vitamin K-dependent coagulation factors decrease in a relatively coordinated manner. However, INR standardization remains conditional. When this biological assumption is not met, clinically relevant non-equivalence may persist (18). To this end, we examine the harmonization of PT/INR from a preanalytical, analytical, and postanalytical perspective, which is primarily structured around a five-tier framework: patient-related biological context and non-parallel factor abnormalities; procedural preanalytical variation; reagent-instrument and endpoint-detection dependence; postanalytical INR parameters, including MNPT determination and ISI verification; and external quality assessment (EQA) commutability and reporting practice.

Preanalytical phase

The preanalytical phase is a major source of coagulation testing variability. Preanalytical errors and unsuitable samples can account for up to 70-77% of all laboratory errors. In coagulation testing specifically, specimen rejection rates are significantly higher than for other test groups (19, 20). This review distinguishes procedural preanalytical variables (directly affecting specimen quality) from patient-related biological conditions (altering the clinical context of PT/INR interpretation). Only procedural variables are cross-institutional standardization targets; biological modifiers define the limits of interpretive comparability (21-23).

Patient-related biological determinants beyond procedural preanalytics

Not all influences preceding PT measurement are procedural. Before venipuncture, the patient’s biological state may already affect both the generation and interpretation of PT prolongation. Unlike classical preanalytical variables, these factors do not primarily impair specimen integrity, but alter the biological context in which PT/INR is produced and interpreted.

This distinction is important because the INR model was developed in VKA-treated patients (24). In other conditions, PT prolongation reflects heterogeneous and often overlapping pathophysiological mechanisms. In chronic liver disease, including cirrhosis, alterations involve not only reduced synthesis of procoagulant factors but also concomitant changes in anticoagulant and fibrinolytic pathways; therefore, PT/INR captures only a limited component of the rebalanced hemostatic system. Moreover, because the INR was standardized primarily for monitoring vitamin K antagonist therapy, it does not reliably harmonize PT results across thromboplastin reagents in patients with chronic liver disease. Inter-reagent differences can reach approximately 0.7 INR units in this setting, underscoring that INR should not be used as a stand-alone measure of global coagulopathy or bleeding risk in cirrhosis (25). In sepsis, by contrast, PT prolongation is more closely related to inflammation-driven coagulation activation, factor consumption, endothelial dysfunction and consumptive coagulopathy, while reduced vitamin K availability in critical illness may provide a partially overlapping and potentially modifiable contribution (26, 27).

From a harmonization perspective, these patient-related determinants may therefore amplify residual differences between reagents and analytical systems even when procedural preanalytics are well controlled (21). They should therefore be considered biological modifiers of PT/INR interpretability rather than conventional preanalytical errors. This distinction is essential, because it explains why optimized collection and processing alone cannot ensure full PT/INR comparability across all clinical settings (3, 8).

Blood collection and phlebotomy-related variables

Proper venipuncture technique is fundamental to the reliability of PT/INR assays. Preanalytical errors occur in roughly 5% of coagulation specimen collections (28). Although the exact proportions reported in different studies differ, insufficient fill volume and clot formation consistently remain the predominant issues (29, 30). In accordance with Clinical and Laboratory Standards Institute (CLSI) guidance, the recommended order of draw positions sodium citrate tubes (light blue, 3.2% citrate) immediately after blood culture bottles and before any serum, heparin, EDTA, or glycolytic inhibitor tubes (31). When a winged (“butterfly”) collection set is used, a discard tube (non-additive or citrate) must be drawn first to offset tubing dead space and avert under-filling; this discard remains mandatory even if only a coagulation tube is collected. No discard is required for standard evacuated tubes without a winged set.

Tourniquet application should not exceed one minute and must be released as soon as blood flow is established (32). Sodium citrate tubes must be filled to at least 90% of the nominal volume to maintain the critical 9:1 blood-to-anticoagulant ratio; under-filling leads to relative citrate excess and consequent falsely prolonged PT/INR values. Studies demonstrate that underfilling 3.8% sodium citrate collection tubes significantly prolongs PT (33). However, when using the 3.2% sodium citrate currently recommended by CLSI, no statistically significant difference in PT results was observed between 60% and 100% fill volumes (31). Immediately after collection, the tube should be gently inverted four to five times to ensure thorough mixing, whereas vigorous shaking should be avoided to prevent hemolysis (34). A clean, minimally traumatic venipuncture minimises tissue factor contamination, and severely hemolyzed samples are considered unacceptable (35). Collectively, strict adherence to these standardised procedures and to the correct order of draw effectively minimises collection-related variability (36).

Anticoagulant ratio, processing, and sample stability

After collection, the in vitro specimen environment is critical. Prothrombin time depends on accurate reversal of citrate-mediated calcium chelation. In markedly elevated hematocrit (> 55%), reduced plasma fraction produces relative anticoagulant excess. Guidelines recommend adjusted citrate volumes, yet the evidence base derives predominantly from indirect studies of underfilled tubes or artificially constructed hematocrit values rather than direct patient data (31). A study demonstrated that the impact of citrate adjustment depends critically on reagent type: recombinant tissue factor thromboplastin showed strong agreement between adjusted and unadjusted INRs (R2 = 0.823; median 2.25 vs 2.22), with > 10% deviation only when hematocrit exceeded 62%, whereas rabbit brain thromboplastin exhibited only moderate correlation (R2 = 0.427) and a clinically significant median INR difference (2.51 vs 3.45), challenging the universal application of current guidelines (37). From a practical standpoint, manual CLSI H21 adjustment is burdensome in high-throughput laboratories, requiring individual patient hematocrit values, manual tube preparation, and careful labeling, each step introducing potential human error, while automation strategies such as pre-labeled adjusted tubes or barcode-integrated calculation systems remain underutilized in most routine settings. Laboratories thus face a trade-off between theoretical accuracy and operational reality.

Prothrombin time/INR stability is time- and temperature-dependent: CLSI recommends up to 24 hours at room temperature, though refrigerated storage data are nuanced (plasma PT/INR is stable for up to 24 hours at 4 °C in some studies; whole blood PT/INR can also be refrigerated for up to 24 hours per the updated guideline, which supersedes previous recommendations against refrigeration (31, 38). For patients receiving unfractionated heparin, whole blood specimens must be centrifuged within 1 hour at room temperature, and plasma analysis completed within 4 hours of collection; delays beyond these thresholds may compromise reliability due to platelet factor 4 mediated heparin neutralization (39). Ultimately, adequate centrifugation, defined as centrifugation of capped citrated whole blood at 1500xg for no less than 15 minutes at room temperature (15-25 °C), preferably using a swing-out bucket rotor with the brake deactivated, to achieve platelet-poor plasma with a residual platelet count < 10,000/μL, and strict adherence to standardized collection, processing, and timing protocols remain essential prerequisites for inter-laboratory comparability (40, 41). Major procedural sources of preanalytical variability affecting PT/INR are shown in Table 1.

Table 1

Major procedural sources of preanalytical variability affecting PT/INR

Variation category Source Mechanism Effect on PT/INR Controllability
Blood collection Prolonged or difficult venipuncture (41) Partial activation during sampling; specimen quality compromise Unreliable PT/INR High
Blood collection Incorrect order of draw in open systems / line contamination (41) Carry-over of anticoagulants or other additives Prolonged or unreliable PT/INR, especially when contamination occurs High
Sample processing Underfilled citrate tube (41,42) Excess citrate ® calcium over-chelation Prolonged PT / elevated INR High
Sample processing Extreme hematocrit (> 55%) without citrate adjustment (42) Relative excess of citrate in reduced plasma volume Prolonged PT / elevated INR Medium
Sample processing Inadequate centrifugation (40,42) Residual cellular components and platelet phospholipid Potentially altered or less reliable PT/INR (direction assay-dependent) High
Transport / storage Delay in processing or inadequate temperature control (42) Time- and condition-dependent alteration of coagulation factor/activity stability Usually prolonged or less reliable PT/INR High
PT/INR - prothrombin time/international normalized ratio. PT – prothrombin time.

Analytical phase

The analytical phase is a critical determinant of PT/INR comparability. Although the ISI framework was introduced to reduce variability attributable to differences in thromboplastin responsiveness, analytical harmonization remains incomplete in routine practice. Residual non-comparability may persist because differences in reagent composition, analyzer-specific clot detection principles, and context-dependent sample behavior are only partially corrected by INR transformation (1-4). Analytical disagreement should therefore be understood not simply as the result of isolated technical imperfections, but as the combined consequence of biological complexity and system-dependent measurement design. This issue is particularly important in multi-center clinical trials, transplant networks involving cross-border organ allocation, or regions with high population diversity. Laboratories should consider local MNPT establishment or formal transfer validation when patient demographics or analytical systems differ substantially from reference conditions (41).

Thromboplastin reagents and analytical sensitivity

The limitations of the ISI model’s foundational assumptions become especially evident outside the classic vitamin VKA setting. The model presupposes coordinated declines in vitamin K-dependent factors, thereby allowing universal calibration correction. When factor abnormalities are non-parallel, the PT prolongation-to-deficiency relationship becomes reagent- and system-dependent. In advanced liver disease, disproportionate factor VII reduction produces reagent-dependent PT responses that cannot be fully corrected by a single ISI conversion (17, 42).

Current guidelines recommend evaluating at least 20 normal plasmas and 60 stable VKA-treated samples spanning INR approximately 1.5-4.5, with log-transformed PT results compared to the reference system via orthogonal regression (43, 44). To address method-dependent bias, Tripodi and colleagues developed INR calibrated for cirrhosis (INR(liver)), using a cirrhosis-specific ISI derived from chronic liver disease patients (45). Independent validation yielded mixed results: Bechmann et al. reported that INR(liver) reduced but did not eliminate method-related Model for End-Stage Liver Disease (MELD) variation, with residual 8-12% disagreement at INR > 2.0 (46). White et al. confirmed improvement was most pronounced at lower INR ranges (1.5-2.5) but diminished with severe factor VII depletion. These data indicate INR(liver) is a partial solution; clinical harmonization requires complementary strategies including ISI optimization and thrombin generation assay (TGA) integration (47, 48). Similar divergence has been documented in sepsis-associated coagulopathy and other non-VKA conditions where factor consumption patterns are heterogeneous (49).

Another issue concerns MNPT sourcing. Laboratories may locally determine MNPT from healthy donors or, in some settings, rely on manufacturer assignments. The latter assumes manufacturer reference conditions match the local reagent-instrument combination, an assumption that may not hold in practice (41, 50). Direct adoption of manufacturer-assigned MNPT may introduce systematic bias if local normal PT values differ. Postanalytical harmonization requires not only correct formula application but validation that the MNPT is suitable for the local measurement system.

Detection principles and endpoint definition

Analytical variability also arises from the way clot formation is detected and translated into a reportable endpoint. Clotting time is assigned through platform-specific detection principles (Table 2). Prothrombin time systems use mechanical, viscosity-based, or optical methods. These differ in susceptibility to interference from hemolysis, lipemia, bilirubin, and hypofibrinogenemia (13).

Table 2

Detection principles in coagulation analyzers and their relevance to PT/INR

Analytical aspect Mechanical/viscosity-based detection Optical detection Selected magnetic bead-based platforms Relevance to PT/INR comparability
Physical principle Detects changes in sample movement, oscillation, or viscosity as fibrin polymerization progresses Detects changes in light transmission, absorbance, or turbidity during clot formation Detects clot-related changes in magnetic particle motion using platform-specific sensing systems Different detection principles may translate the same coagulation reaction into different PT seconds, which are not always fully normalized by INR conversion
Endpoint definition Endpoint is assigned when movement attenuation or viscosity-related threshold criteria are reached Endpoint is assigned from optical curve thresholds and platform-specific signal-processing algorithms Endpoint is assigned from changes in magnetic particle behavior together with platform-specific threshold algorithms Endpoint assignment is operational rather than universal, and may therefore contribute to inter-system variability
Susceptibility to hemolysis/lipemia/icterus Less susceptible to purely optical interference, although biological and matrix-related effects may still influence results More susceptible to optical interference, although modern HIL checks and correction algorithms may mitigate some effects In some comparative studies, showed lower susceptibility to optical interference than photo-optical systems; performance remains platform-dependent Matrix-dependent bias differs across systems and may reduce comparability in visually compromised samples
Performance in low fibrinogen states Weak clot formation may still impair endpoint recognition, particularly in severe hypofibrinogenemia Weak or shallow optical clotting curves may reduce signal robustness Performance may remain acceptable on some platforms, but depends on assay design and signal-processing rules Severe hypofibrinogenemia can accentuate system-dependent differences in clot endpoint assignment
Additional analytical output Usually limited to endpoint-based clotting results May provide full clotting curves and derived waveform parameters; clot waveform analysis is primarily associated with photo-optical systems Usually limited in routine PT output, depending on platform implementation Additional algorithmic output may improve interpretation, but also increases platform dependence
Practical implication Often advantageous in optically compromised samples, but not inherently interchangeable with other systems Informative and waveform-capable, but more sensitive to optical artifacts Potentially useful in selected optically compromised specimens, but not inherently interchangeable with other systems PT/INR harmonization should evaluate the reagent–instrument pair as the analytical unit rather than the reagent alone
PT/INR - prothrombin time/international normalized ratio. HIL - hemolysis/icterus/lipemia index. PT – prothrombin time.

Traditional internal quality control (IQC) and EQA remain laboratory quality assurance foundations. However, satisfactory performance in traditional quality systems does not guarantee clinically equivalent results across laboratories, because PT/INR comparability depends not only on analytical precision but also on whether quality assessment materials behave like real patient plasma across different reagent-instrument systems (51).

Analytical non-comparability is therefore not reducible to reagent sensitivity alone. Instead, PT/INR performance should be understood at the level of the integrated analyzer–reagent system, in which the same biological reaction may be translated into different clotting times depending on how endpoint formation is detected and processed. For PT/INR harmonization, the relevant unit of comparison is thus not the reagent in isolation, but the full measurement system through which coagulation is measured and converted into reportable data (see Table 2).

Pharmacodynamic interference from direct oral anticoagulants

The increasing use of DOACs illustrates an important boundary of PT/INR harmonization outside the VKA model. Unlike VKAs, which reduce several vitamin K-dependent clotting factors in a relatively coordinated manner, DOACs exert target-specific effects on coagulation assays (14). Their influence on PT therefore reflects drug-specific and reagent-dependent analytical effects rather than a biological pattern that can be reliably normalized through conventional INR transformation. A systematic review by Samuelson et al. demonstrated that the performance of standard anticoagulation tests varies across DOACs and reagents, with most assays showing insufficient correlation to provide a reliable assessment of DOAC effects (16).

However, recent developments include the MRX PT DOAC assay, which generates a DOAC Normalised Ratio (DNR) from a DOAC-sensitive PT and an INR from a DOAC-insensitive PT, with the Clot Time Ratio (CTR) calculated as DNR/INR (52). In a 2024 evaluation, the MRX PT DOAC assay demonstrated 100% sensitivity and 100% negative predictive value for edoxaban concentrations > 50 ng/mL, and can be run on widely available coagulation analyzers (53). This suggests that reagent-dependent DOAC effects may be harnessed for detection purposes rather than dismissed solely as confounding variability. Direct oral anticoagulant stop reagents, such as DOAC Stop and DP-Filter, offer another pathway: these products can effectively neutralize the in vitro effect of all DOACs, including rivaroxaban, enabling accurate baseline PT/INR determination in DOAC-exposed samples (54, 55). Future guidelines may evolve to incorporate DOAC-specific verification protocols, particularly for emergency settings where rapid exclusion of clinically relevant DOAC concentrations is required. Typical effects of selected DOACs on PT assays and associated interpretation risks are shown in Table 3.

Table 3

Typical effects of selected DOACs on PT assays and associated interpretation risks

Drug class/example Typical effect on PT Inter-reagent variability Main interpretation risk Recommended reporting implication References
Direct factor Xa inhibitor (rivaroxaban) Often moderate to marked PT prolongation High PT prolongation may be misread as a standardized estimate of anticoagulant intensity PT/INR should not be used to estimate rivaroxaban intensity; if assessment is clinically required, use a drug-appropriate assay ((14,15,56))
Direct factor Xa inhibitor (apixaban) Usually minimal, absent, or inconsistent effect in many PT systems High A normal PT may falsely reassure clinicians despite clinically relevant drug exposure A normal PT does not exclude clinically relevant apixaban exposure ((14,15,57))
Direct thrombin inhibitor (dabigatran) Usually weak and inconsistent PT effect; prolongation is more likely at higher concentrations Moderate to high PT is relatively insensitive and may underestimate drug presence PT is not suitable for excluding dabigatran; dilute thrombin time or ecarin-based assays are preferred when assessment is required ((14,56,57))
Direct factor Xa inhibitor (edoxaban) Moderate PT prolongation, greater than apixaban but less than rivaroxaban in most systems (1.5-2.5 x control at peak) Moderate to high INR may appear falsely elevated; edoxaban-specific anti-Xa assay preferred for quantitative assessment PT/INR should not be used to estimate edoxaban intensity; anti-Xa assay recommended if quantitative data required ((14,56))
DOACs overall Drug-specific and reagent-dependent PT effects that do not conform to the VKA calibration model High INR reporting may give a misleading impression of standardization and comparability PT results in DOAC-treated patients should not be interpreted within the conventional VKA-oriented INR framework ((14,56,57))
DOACs - direct oral anticoagulants. PT – prothrombin time. PT/INR - prothrombin time/international normalized ratio. VKA - vitamin K antagonist.

Artificial intelligence and machine learning in PT/INR quality assurance

Artificial intelligence (AI) and machine learning (ML) approaches have recently been explored in laboratory quality systems, particularly for patient-based real-time quality control (PBRTQC) and coagulation waveform analysis (58-60). Existing studies using ensemble learning and rule-based hybrid models suggest that ML may improve the early detection of analytical drift; however, the magnitude of improvement is strongly dependent on dataset structure, local patient mix, and calibration strategy, and cannot be generalized as a fixed performance gain across laboratories (58). For example, reported improvements in drift detection sensitivity vary widely across simulation-based and retrospective datasets, and are influenced by baseline analytical stability and signal-to-noise ratio rather than algorithm type alone (58, 59). Consequently, performance estimates should be interpreted as context-specific rather than transferable metrics, and prospective multicenter validation remains limited.

Deep learning models applied to optical clot waveform data have demonstrated promising capability in identifying preanalytical interference such as hemolysis, lipemia, and icterus. Nevertheless, most published studies remain retrospective or single-platform in design, and clinical implementation requires validation across heterogeneous analyzer-reagent systems before routine adoption (61). Importantly, current evidence supports AI as an adjunctive interpretive layer rather than a replacement for standardized analytical quality systems. Integration into routine laboratory workflows will require resolution of key barriers, including model transparency, regulatory validation, interoperability with laboratory information systems, and dataset harmonization across institutions (60, 62).

Point-of-care testing INR devices and clinical harmonization implication

The expansion of point-of-care testing (POCT) has extended the PT/INR harmonization challenge beyond centralized laboratories into decentralized clinical settings (63). While POCT INR devices improve care accessibility and enable rapid feedback for outpatient and home VKA monitoring, they cannot be assumed analytically interchangeable with central laboratory assays without formal verification (64). Point-of-care testing systems adopt strip- or cartridge-based whole-blood detection, whereas laboratory assays use platelet-poor plasma with optical or mechanical clot detection, creating inherent gaps in detection principle and matrix susceptibility (65, 66).

Agreement between the two methods is generally acceptable within conventional therapeutic INR ranges, but bias progressively increases at higher values, driven by analytical non-linearity and intrinsic matrix effects of whole-blood testing (67). Discordance risk is further amplified in patients with advanced liver disease or abnormal hematocrit and fibrinogen concentrations. This carries clinical relevance: MELD scores are standardized to laboratory INR, while POCT results commonly seen in emergency or ambulatory settings may lead to misclassification of disease severity (63, 68).

Regular comparative verification against local central laboratory systems is essential. Current guidance recommends method comparison in stable VKA-treated patients, with an acceptable paired deviation of approximately 0.5 INR units (69). Venous laboratory confirmation should be prioritized before major clinical decisions at high INR levels (65). U.S. Food and Drug Administration (FDA) recalls of INRatio devices further underscore the necessity of device-specific validation, even as newer platforms show improved performance (67, 70). Point-of-care testing does not undermine the broader INR framework, but clarifies that cross-system comparability remains conditional on measurement method, specimen matrix, and clinical context.

Owren-Quick Divergence and Fiix-PT Calibration

The persistent methodological divergence between Owren and Quick PT systems remains a relevant source of inter-laboratory variability despite shared ISI-based calibration (71). Although INR transformation improves comparability, residual bias has been consistently demonstrated at higher INR ranges and in complex patient populations, particularly in the context of liver disease and intensive anticoagulation regimens (68).

The Fiix-PT concept offers an innovative approach to reducing analytical variability by selectively prioritizing factor II and X activity while attenuating sensitivity to factor VII-driven fluctuations (72). This strategy may enhance analytical stability in VKA monitoring; however, its clinical adoption remains at an early validation stage, with current evidence largely derived from observational and regionally confined studies (73). Notably, while the Icelandic nationwide cohort study demonstrated improved VKA stability and reduced bleeding events, these findings were obtained within a relatively homogeneous healthcare system with centralized laboratory standardization, which may limit their direct extrapolation to more heterogeneous international settings (74).

Accordingly, Fiix-PT should currently be regarded as a promising yet context-sensitive calibration strategy, requiring robust multicenter validation across diverse reagent–instrument systems before broad implementation can be recommended (73). From a harmonization perspective, it is more appropriately viewed as part of a broader conceptual evolution toward mechanism-informed coagulation assessment rather than as a direct substitute for conventional PT/INR. Future research should prioritize: (i) multicenter validation across different reagent classes; (ii) head-to-head comparisons with standardized INR in diverse populations; and (iii) integration into clinical decision algorithms, rather than relying on isolated numerical substitution.

Postanalytical phase

Postanalytical variability in PT/INR testing arises mainly from the parameters used to convert PT seconds into INR, the compatibility of those parameters with the local reagent–instrument system, and the way INR results are subsequently interpreted or incorporated into clinical decision tools (43). Although these factors occur after clot detection, they can introduce clinically relevant bias because INR is a calculated result rather than a directly measured analyte.

Determination of the mean normal prothrombin time

Mean normal prothrombin time serves as the baseline denominator in the INR formula, and systematic bias in this value propagates directly into all patient INR results (43). Manufacturers typically provide a default MNPT value established under reference conditions, but this value may not transfer reliably to local reagent-instrument combinations or patient populations. Population-related variation in normal PT values represents an often-overlooked source of systematic bias: reference intervals for coagulation tests established in Western cohorts may not be appropriate for non-Western populations, and direct adoption of imported MNPT values can introduce consistent offset across all patient results (75).

Current CLSI guidelines recommend that laboratories establish or verify MNPT locally using at least 20 plasma samples from healthy donors matched to the local patient demographic (43). For open analytical systems where reagents are used on non-validated instrument platforms, local MNPT determination is not optional but a methodological requirement to maintain metrological traceability. Field evidence indicates that laboratories performing local MNPT verification exhibit significantly lower between-laboratory INR bias than those relying exclusively on manufacturer-assigned values (76). In multinational clinical trials or cross-border organ allocation networks, formal transfer validation of MNPT is particularly critical, as even small deviations in baseline PT can produce clinically meaningful INR differences near therapeutic or prognostic thresholds.

International Sensitivity Index verification and postanalytical system compatibility

The ISI is the core calibration parameter that normalizes thromboplastin sensitivity differences across reagents. However, manufacturer-assigned ISI values are validated only for specific reagent-instrument combinations; when reagents are deployed on alternative platforms in open systems, the ISI may no longer be accurate, introducing systematic bias into INR calculations (3, 77).

Per CLSI H54-A guidelines, local ISI calibration or verification requires testing at least 20 normal plasmas and 60 stable VKA-treated samples spanning an INR range of approximately 1.5-4.5, with orthogonal regression applied to log-transformed PT results against a reference measurement system (43). This procedure corrects for platform-specific differences in clot endpoint detection, optical signal processing, and reagent-instrument interaction that cannot be captured by a generic reagent ISI. Multi-center evaluation data demonstrate that local ISI calibration reduces between-laboratory INR bias compared with using manufacturer-assigned values alone, particularly at higher INR values (78). Despite this well-documented benefit, local calibration remains underimplemented in routine practice, contributing to persistent postanalytical variability across institutions.

Postanalytical harmonization therefore requires recognition that INR is generated by an integrated reagent-instrument-calculation system, not by reagent sensitivity alone. Regulatory frameworks increasingly emphasize that manufacturers should provide system-specific ISI values for each certified reagent-instrument pair, rather than a single generic reagent ISI, to reduce the burden on individual laboratories and improve baseline standardization.

Clinical amplification of INR variability

The clinical impact of postanalytical INR variability is disproportionately magnified when INR values are incorporated into validated prognostic scoring systems, most notably the MELD score (68). The MELD algorithm uses a log-transformed INR term, meaning that even small absolute differences in INR (0.2-0.3 units) can alter calculated disease severity and shift patient priority near liver transplantation eligibility thresholds (79).

This amplification effect is particularly consequential in advanced liver disease, where INR itself is already subject to substantial reagent-dependent analytical bias due to non-parallel factor deficiency (25). Postanalytical errors in MNPT or ISI thus compound pre-existing analytical non-equivalence, leading to clinically significant misclassification of cirrhosis severity. To minimize this cumulative variation, laboratories serving liver transplant centers are advised to prioritize thromboplastin reagents with ISI values closest to 1.0, implement local ISI/MNPT calibration, and participate in EQA programs using commutable human plasma materials (25, 80).

Future outlooks

Although the INR system has substantially improved comparability of PT results for monitoring vitamin K antagonist therapy, complete harmonization across laboratories remains difficult (5, 8). Future research should focus on four areas: local verification of MNPT and ISI in open reagent-instrument systems; broader use of commutable EQA materials; indication-specific interpretation for clinical settings in which conventional INR assumptions are not fulfilled; and prospective validation of complementary approaches such as Fiix-PT, thrombin generation assays, PBRTQC, and AI-assisted quality surveillance. These priorities require multicentre studies including diverse patient populations, different thromboplastin reagents, Quick and Owren methods, central-laboratory and decentralized testing environments, and clinically relevant outcomes.

Conclusion

The INR remains one of the most influential standardization tools in laboratory medicine, but its success should not obscure its limitations. As this review shows, calibration of thromboplastin responsiveness alone does not ensure full PT comparability across the total testing process or across all clinical settings. Residual variability may arise from patient-related biological context, specimen handling, reagent-instrument interactions, postanalytical parameter validity, and decentralized testing conditions. Prothrombin time/INR harmonization should therefore be understood not as a completed achievement, but as an ongoing process requiring integrated control across preanalytical, analytical, and postanalytical phases together with quality systems that reflect real-world patient samples and routine practice.

Acknowledgments

This work was supported by the Luzhou Medical Association (Grant No. 2024-YXXM-120). During the preparation of this manuscript, the authors used Kimi (Moonshot AI) for the purposes of polishing the language and style. The authors have reviewed and edited the output and take full responsibility for the content of this publication.

Notes

[1] Conflicts of interest Potential conflict of interest

None declared.

Data availability statement

No data was generated during this study, so data sharing statement is not applicable to this article.

References

1 

Gosselin RC. Review of coagulation preanalytical variables with update on the effect of direct oral anticoagulants. Int J Lab Hematol. 2021;43 Suppl 1:109–16. https://doi.org/10.1111/ijlh.13585

2 

Kitchen S, Adcock DM, Dauer R, Kristoffersen AH, Lippi G, Mackie I, et al. International Council for Standardization in Haematology (ICSH) recommendations for processing of blood samples for coagulation testing. Int J Lab Hematol. 2021;43:1272–83. https://doi.org/10.1111/ijlh.13702

3 

Poller L. International Normalized Ratios (INR): the first 20 years. J Thromb Haemost. 2004;2:849–60. https://doi.org/10.1111/j.1538-7836.2004.00775.x

4 

Dorgalaleh A, Favaloro EJ, Bahraini M, Rad F. Standardization of Prothrombin Time/International Normalized Ratio (PT/INR). Int J Lab Hematol. 2021;43:21–8. https://doi.org/10.1111/ijlh.13349

5 

van den Besselaar AM, Stavelin A, Kitchen S, Bryant M, Tripodi A, Scalambrino E, et al. Defining a metrologically traceable and sustainable calibration hierarchy of international normalized ratio for monitoring of vitamin K antagonist treatment in accordance with International Organization for Standardization (ISO) 17511:2020 standard: communication from the International Federation of Clinical Chemistry and Laboratory Medicine-SSC/ISTH working group on prothrombin time/international normalized ratio standardization. J Thromb Haemost. 2024;22:1236–48. https://doi.org/10.1016/j.jtha.2023.12.013

6 

van den Besselaar AM, van Rijn CJJ, Abdoel CF, Chantarangkul V, Scalambrino E, Kitchen S, et al. Paving the way for establishing a reference measurement system for standardization of plasma prothrombin time: Harmonizing the manual tilt tube method. J Thromb Haemost. 2020;18:1986–94. https://doi.org/10.1111/jth.14873

7 

Kirkwood TB. Calibration of reference thromboplastins and standardisation of the prothrombin time ratio. Thromb Haemost. 1983;49:238–44. https://doi.org/10.1055/s-0038-1657371

8 

Plebani M, Lippi G. Standardization and harmonization in laboratory medicine: not only for clinical chemistry measurands. Clin Chem Lab Med. 2022;61:185–7. https://doi.org/10.1515/cclm-2022-1122

9 

Bai B, Christie DJ, Gorman RT, Wu JR. Comparison of optical and mechanical clot detection for routine coagulation testing in a large volume clinical laboratory. Blood Coagul Fibrinolysis. 2008;19:569–76. https://doi.org/10.1097/MBC.0b013e3283070872

10 

Tekkesin N, Kılınc C. Optical and mechanical clot detection methodologies: a comparison study for routine coagulation testing. J Clin Lab Anal. 2012;26:125–9. https://doi.org/10.1002/jcla.21497

11 

Geens T, Vertessen F, Malfait R, Deiteren K, Maes MB. Validation of the Sysmex CS5100 coagulation analyzer and comparison to the Stago STA-R analyzer for routine coagulation parameters. Int J Lab Hematol. 2015;37:372–81. https://doi.org/10.1111/ijlh.12295

12 

Ülfer G. A comparison of the optical method with the mechanical method in routine coagulation tests. J Med Biochem. 2025;44:479–85. https://doi.org/10.5937/jomb0-56100

13 

Woolley A, Golmard JL, Kitchen S. Effects of haemolysis, icterus and lipaemia on coagulation tests as performed on Stago STA-Compact-Max analyser. Int J Lab Hematol. 2016;38:375–88. https://doi.org/10.1111/ijlh.12498

14 

Douxfils J, Ageno W, Samama CM, Lessire S, Ten Cate H, Verhamme P, et al. Laboratory testing in patients treated with direct oral anticoagulants: a practical guide for clinicians. J Thromb Haemost. 2018;16:209–19. https://doi.org/10.1111/jth.13912

15 

Gosselin RC, Adcock DM, Bates SM, Douxfils J, Favaloro EJ, Gouin-Thibault I, et al. International Council for Standardization in Haematology (ICSH) Recommendations for Laboratory Measurement of Direct Oral Anticoagulants. Thromb Haemost. 2018;118:437–50. https://doi.org/10.1055/s-0038-1627480

16 

Samuelson BT, Cuker A, Siegal DM, Crowther M, Garcia DA. Laboratory Assessment of the Anticoagulant Activity of Direct Oral Anticoagulants: A Systematic Review. Chest. 2017;151:127–38. https://doi.org/10.1016/j.chest.2016.08.1462

17 

Lisman T, van Leeuwen Y, Adelmeijer J, Pereboom IT, Haagsma EB, van den Berg AP, et al. Interlaboratory variability in assessment of the model of end-stage liver disease score. Liver Int. 2008;28:1344–51. https://doi.org/10.1111/j.1478-3231.2008.01783.x

18 

Winter WE, Flax SD, Harris NS. Coagulation Testing in the Core Laboratory. Lab Med. 2017;48:295–313. https://doi.org/10.1093/labmed/lmx050

19 

Aquino JB, Damian KB. Pre-analytical Factors Influencing Blood Sample Rejection Rate in the Hematology Laboratory of the Philippine General Hospital from 2018 to 2022: A Cross-sectional Study. Acta Med Philipp. 2025;59:30–6.

20 

Goswami B, Mahajan B, Chawla R, Mallika V. Evaluation of errors in a clinical laboratory: A one-year experience. Clin Chem Lab Med. 2010;48:63–6. https://doi.org/10.1515/CCLM.2010.006

21 

Lisman T, Hernandez-Gea V, Magnusson M, Roberts L, Stanworth S, Thachil J, et al. The concept of rebalanced hemostasis in patients with liver disease: Communication from the ISTH SSC working group on hemostatic management of patients with liver disease. J Thromb Haemost. 2021;19:1116–22. https://doi.org/10.1111/jth.15239

22 

Dahlberg S, Schött U, Kander T. The effect of vitamin K on prothrombin time in critically ill patients: an observational registry study. J Intensive Care. 2021;9:11. https://doi.org/10.1186/s40560-020-00517-5

23 

Iba T, Levi M, Levy JH. Sepsis-Induced Coagulopathy and Disseminated Intravascular Coagulation. Semin Thromb Hemost. 2020;46:89–95. https://doi.org/10.1055/s-0039-1694995

24 

Zivelin A, Rao LV, Rapaport SI. Mechanism of the anticoagulant effect of warfarin as evaluated in rabbits by selective depression of individual procoagulant vitamin K-dependent clotting factors. J Clin Invest. 1993;92:2131–40. https://doi.org/10.1172/JCI116814

25 

Tripodi A, Chantarangkul V, Mannucci PM. The international normalized ratio to prioritize patients for liver transplantation: problems and possible solutions. J Thromb Haemost. 2008;6:243–8. https://doi.org/10.1111/j.1538-7933.2008.02827.x

26 

van Vught LA, Uhel F, Ding C, Van’t Veer C, Scicluna BP, Peters-Sengers H, et al. Consumptive coagulopathy is associated with a disturbed host response in patients with sepsis. J Thromb Haemost. 2021;19:1049–63. https://doi.org/10.1111/jth.15246

27 

Paulus MC, Drent M, Kouw IWK, Balvers MGJ, Bast A, van Zanten ARH. Vitamin K: a potential missing link in critical illness-a scoping review. Crit Care. 2024;28:212. https://doi.org/10.1186/s13054-024-05001-2

28 

Kitchen S, Adcock DM, Dauer R, Kristoffersen AH, Lippi G, Mackie I, et al. International Council for Standardisation in Haematology (ICSH) recommendations for collection of blood samples for coagulation testing. Int J Lab Hematol. 2021;43:571–80. https://doi.org/10.1111/ijlh.13584

29 

Gaur K, Puri V, Shukla S, Sharma S, Suman S, Singh R. Finish before the start: Analyzing preanalytical sample errors in a tertiary care hematology laboratory. Indian J Pathol Microbiol. 2020;63:435–40. https://doi.org/10.4103/IJPM.IJPM_191_19

30 

Iqbal MS, Tabassum A, Arbaeen AF, Qasem AH, Elshemi AG, Almasmoum H. Preanalytical Errors in a Hematology Laboratory: An Experience from a Tertiary Care Center. Diagnostics (Basel). 2023;13: https://doi.org/10.3390/diagnostics13040591

31 

Clinical and Laboratory Standards Institute. Collection, Transport, and Processing of Blood Specimens for Testing Plasma-Based Coagulation Assays. 6th ed. CLSI guideline H21. Wayne, PA: Clinical and Laboratory Standards Institute; 2024.

32 

Gosselin RC, Marlar RA. Preanalytical Variables in Coagulation Testing: Setting the Stage for Accurate Results. Semin Thromb Hemost. 2019;45:433–48. https://doi.org/10.1055/s-0039-1692700

33 

Kim HK, Park S, Lim JH, Jeong J, Lee S. Validation of pre-analytical stability of specimens requested for various routine coagulation tests. Clin Exp Thromb Hemost. 2021;7:7–13. https://doi.org/10.14345/ceth.21002

34 

Adcock Funk DM, Lippi G, Favaloro EJ. Quality standards for sample processing, transportation, and storage in hemostasis testing. Semin Thromb Hemost. 2012;38:576–85. https://doi.org/10.1055/s-0032-1319768

35 

Baker P, Platton S, Gibson C, Gray E, Jennings I, Murphy P, et al. Guidelines on the laboratory aspects of assays used in haemostasis and thrombosis. Br J Haematol. 2020;191:347–62. https://doi.org/10.1111/bjh.16776

36 

Plebani M, Sciacovelli L, Aita A, Padoan A, Chiozza ML. Quality indicators to detect pre-analytical errors in laboratory testing. Clin Chim Acta. 2014;432:44–8. https://doi.org/10.1016/j.cca.2013.07.033

37 

Silva VM, Rezende DC, Garcia ES, Cavalheiro C, Strunz CC. Effect of anticoagulant adjustment on prothrombin time test using two different PT reagents in patients with elevated hematocrit. Pract Lab Med. 2020;22:e00177. https://doi.org/10.1016/j.plabm.2020.e00177

38 

Nakanishi R, Sakayori T, Matsui D, Kono M, Maki M, Dunois C, et al. Effects of storage time and temperature on coagulation factor and natural anticoagulant activities in healthy individuals. Sci Rep. 2025;15:10797. https://doi.org/10.1038/s41598-025-95389-w

39 

Baker P, Platton S, Arachchillage DJ, Kitchen S, Patel J, Riat R, et al. Measurement of heparin, direct oral anti-coagulants and other non-coumarin anti-coagulants and their effects on haemostasis assays: A British Society for Haematology Guideline. Br J Haematol. 2024;205:1302–18. https://doi.org/10.1111/bjh.19729

40 

Gosselin RC. Ultracentrifugation for Coagulation Testing. Methods Mol Biol. 2023;2663:63–70. https://doi.org/10.1007/978-1-0716-3175-1_4

41 

Favaloro EJ, Adcock DM. Standardization of the INR: how good is your laboratory’s INR and can it be improved? Semin Thromb Hemost. 2008;34:593–603. https://doi.org/10.1055/s-0028-1104538

42 

Lee JH, Kweon OJ, Lee MK, Lee HW, Kim HJ, Kim HR. Clinical usefulness of international normalized ratio calibration of prothrombin time in patients with chronic liver disease. Int J Hematol. 2015;102:163–9. https://doi.org/10.1007/s12185-015-1820-2

43 

Adcock DM, Brien WF, Duff SL, Johnston M, Marlar RA, Ng VL, et al. Procedures for validation of INR and local calibration of PT/INR systems; approved guideline. CLSI document H54-A. Wayne, PA: Clinical and Laboratory Standards Institute; 2005.

44 

World Health Organization Expert Committee on Biological Standardization. Guidelines for thromboplastins and plasma used to control oral anticoagulant therapy with vitamin K antagonists. WHO Technical Report Series. 2013;979:271–316.

45 

Tripodi A, Chantarangkul V, Primignani M, Fabris F, Dell’Era A, Sei C, et al. The international normalized ratio calibrated for cirrhosis (INR(liver)) normalizes prothrombin time results for model for end-stage liver disease calculation. Hepatology. 2007;46:520–7. https://doi.org/10.1002/hep.21732

46 

Bechmann LP, Sichau M, Wichert M, Gerken G, Kröger K, Hilgard P. Low-molecular-weight heparin in patients with advanced cirrhosis. Liver Int. 2011;31:75–82. https://doi.org/10.1111/j.1478-3231.2010.02358.x

47 

Hemker HC, Giesen P, Al Dieri R, Regnault V, de Smedt E, Wagenvoord R, et al. Calibrated automated thrombin generation measurement in clotting plasma. Pathophysiol Haemost Thromb. 2003;33:4–15. https://doi.org/10.1159/000071636

48 

Tripodi A, Salerno F, Chantarangkul V, Clerici M, Cazzaniga M, Primignani M, et al. Evidence of normal thrombin generation in cirrhosis despite abnormal conventional coagulation tests. Hepatology. 2005;41:553–8. https://doi.org/10.1002/hep.20569

49 

Taylor FBJr, Toh CH, Hoots WK, Wada H, Levi M. Towards definition, clinical and laboratory criteria, and a scoring system for disseminated intravascular coagulation. Thromb Haemost. 2001;86:1327–30. https://doi.org/10.1055/s-0037-1616068

50 

Critchfield GC, Bennett ST. The influence of the reference mean prothrombin time on the international normalized ratio. Am J Clin Pathol. 1994;102:806–11. https://doi.org/10.1093/ajcp/102.6.806

51 

International Organization for Standardization. ISO 15189:2022. Medical laboratories - Requirements for quality and competence. 4th ed. Geneva: International Organization for Standardization; 2022.

52 

Favaloro EJ, Pasalic L. Innovative Diagnostic Solutions in Hemostasis. Diagnostics (Basel). 2024;14:2521. https://doi.org/10.3390/diagnostics14222521

53 

Mithoowani S, Siegal D. DOACs: role of anti-Xa and drug level monitoring. Hematology (Am Soc Hematol Educ Program). 2024;2024:178–85. https://doi.org/10.1182/hematology.2024000666

54 

Wang J, Lim HY, Nandurkar H, Ho P. DOAC-stop can remove direct oral anticoagulants and allow analysis by global coagulation assays. Int J Lab Hematol. 2023;45:360–7. https://doi.org/10.1111/ijlh.14020

55 

Savola P, Lemponen M, Joutsi-Korhonen L, Helin TA. Novel Ex Vivo DOAC Removal Methods Reduce Interference in Lupus Anticoagulant Testing. Diagnostics (Basel). 2022;12:2520. https://doi.org/10.3390/diagnostics12102520

56 

Douxfils J, Adcock DM, Bates SM, Favaloro EJ, Gouin-Thibault I, Guillermo C, et al. 2021 update of the International Council for Standardization in Haematology recommendations for laboratory measurement of direct oral anticoagulants. Thromb Haemost. 2021;121:1008–20. https://doi.org/10.1055/a-1450-8178

57 

Steffel J, Collins R, Antz M, Cornu P, Desteghe L, Haeusler KG, et al. 2021 European Heart Rhythm Association Practical Guide on the Use of Non-Vitamin K Antagonist Oral Anticoagulants in Patients with Atrial Fibrillation. Europace. 2021;23:1612–76. https://doi.org/10.1093/europace/euab065

58 

Lorde N, Mahapatra S, Kalaria T. Machine Learning for Patient-Based Real-Time Quality Control (PBRTQC), Analytical and Preanalytical Error Detection in Clinical Laboratory. Diagnostics (Basel). 2024;14:1808. https://doi.org/10.3390/diagnostics14161808

59 

Zhou B, Li X, Cheng S, Zhou Z, Kang H. Patient-based pre-classified real-time quality control with neural network (PCRTQC-NN). Pract Lab Med. 2025;47:e00506. https://doi.org/10.1016/j.plabm.2025.e00506

60 

Yang HS, Çubukçu HC, Del Ben F, Duan X, Feng J, Frans G, et al. Implementation of AI systems in the clinical laboratory: insights from an expert survey and recommendations for best practice. Clin Chem Lab Med. 2026;64:1513–26. https://doi.org/10.1515/cclm-2025-1322

61 

Tan JY, De Guzman MRT, Wong WH, Yeo CK, Goh GH, Ng HJ, et al. Decoding Clot Waveform Analysis: Toward Better Understanding and Harmonization. Semin Thromb Hemost. 2026;52:460–75. https://doi.org/10.1055/a-2778-9810

62 

Çubukçu HC, Boursier G, Linko S, Bernabeu-Andreu FA, Meško Brguljan P, Tosheska-Trajkovska K, et al. Regulating the future of laboratory medicine: European regulatory landscape of AI-driven medical device software in laboratory medicine. Clin Chem Lab Med. 2025;63:1891–914. https://doi.org/10.1515/cclm-2025-0482

63 

Kardjadj M. Regulatory Approved Point-of-Care Diagnostics (FDA & Health Canada): A Comprehensive Framework for Analytical Validity, Clinical Validity, and Clinical Utility in Medical Devices. J Appl Lab Med. 2025;10:1622–37. https://doi.org/10.1093/jalm/jfaf106

64 

Wool GD. Benefits and Pitfalls of Point-of-Care Coagulation Testing for Anticoagulation Management: An ACLPS Critical Review. Am J Clin Pathol. 2019;151:1–17. https://doi.org/10.1093/ajcp/aqy087

65 

Clinical and Laboratory Standards Institute. Point-of-Care Coagulation Testing and Anticoagulation Monitoring. 2nd ed. CLSI guideline POCT14. Wayne, PA: Clinical and Laboratory Standards Institute; 2020.

66 

Fitzmaurice DA, Geersing GJ, Armoiry X, Machin S, Kitchen S, Mackie I. ICSH guidance for INR and D-dimer testing using point of care testing in primary care. Int J Lab Hematol. 2023;45:276–81. https://doi.org/10.1111/ijlh.14051

67 

Munir R, Schapkaitz E, Noble L, Loonat S, McCree M, Ali N, et al. A Comprehensive Clinical Assessment of the LumiraDx International Normalized Ratio (INR) Assay for Point-of-Care Monitoring in Anticoagulation Therapy. Diagnostics (Basel). 2024;14:2683. https://doi.org/10.3390/diagnostics14232683

68 

Shikdar S, Vashisht R, Zubair M, Bhattacharya PT. International Normalized Ratio: Assessment, Monitoring, and Clinical Implications. StatPearls. Treasure Island (FL): StatPearls Publishing; 2026.

69 

Bhat M, Abuzied Y, Fagih Z, Wani T, Gawan I, Andalay EM, et al. Efficacy of Point-of-Care for INR Testing Compared to Standard Laboratory Methods at a Tertiary Care Hospital in Saudi Arabia. Glob J Qual Saf Healthc. 2020;3:98–104. https://doi.org/10.36401/JQSH-19-36

70 

U.S. Food and Drug Administration. Class 1 Device Recall: INRatio/INRatio2 PT/INR Test Strips. Silver Spring, MD: U.S. Food and Drug Administration; 2014. Available from: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRES/res.cfm?id=126845. Accessed July 3, 2026.

71 

Nilsson CU, Strandberg K, Reinstrup P. Warfarin monitoring with viscoelastic haemostatic assays, thrombin generation, coagulation factors and correlations to Owren and Quick prothrombin time. Scand J Clin Lab Invest. 2018;78:358–64. https://doi.org/10.1080/00365513.2018.1474492

72 

Jonsson PI, Letertre L, Juliusson SJ, Gudmundsdottir BR, Francis CW, Onundarson PT. During warfarin induction, the Fiix-prothrombin time reflects the anticoagulation level better than the standard prothrombin time. J Thromb Haemost. 2017;15:131–9. https://doi.org/10.1111/jth.13549

73 

Onundarson PT, Palsson R, Witt DM, Gudmundsdottir BR. Replacement of traditional prothrombin time monitoring with the new Fiix prothrombin time increases the efficacy of warfarin without increasing bleeding. A review article. Thromb J. 2021;19:72. https://doi.org/10.1186/s12959-021-00327-1

74 

Oskarsdottir AR, Gudmundsdottir BR, Jensdottir HM, Flygenring B, Palsson R, Onundarson PT. Ignoring instead of chasing after coagulation factor VII during warfarin management: an interrupted time series study. Blood. 2021;137:2745–55. https://doi.org/10.1182/blood.2020008698

75 

Kelem A, Engidaye G, Addisu B, Mandefro B, Teketelew BB, Berta DM, et al. Coagulation parameters reference interval for adult population of Debre Berhan town, Northeast Ethiopia. Thromb J. 2025;23:115. https://doi.org/10.1186/s12959-025-00808-7

76 

Poller L, Keown M, Ibrahim S, van den Besselaar AM, Roberts C, Stevenson K, et al. Comparison of local International Sensitivity Index calibration and ‘Direct INR’ methods in correction of locally reported International Normalized Ratios: an international study. J Thromb Haemost. 2007;5:1002–9. https://doi.org/10.1111/j.1538-7836.2007.02452.x

77 

Favaloro EJ, McVicker W, Lay M, Ahuja M, Zhang Y, Hamdam S, et al. Harmonizing the International Normalized Ratio (INR): standardization of methods and use of novel strategies to reduce interlaboratory variation and bias. Am J Clin Pathol. 2016;145:191–202. https://doi.org/10.1093/ajcp/aqv022

78 

Poller L, Ibrahim S, Keown M, Pattison A, Jespersen J. Simplified method for international normalized ratio (INR) derivation based on the prothrombin time/INR line: an international study. Clin Chem. 2010;56:1608–17. https://doi.org/10.1373/clinchem.2009.141937

79 

Kim WR, Mannalithara A, Heimbach JK, Kamath PS, Asrani SK, Biggins SW, et al. MELD 3.0: The Model for End-Stage Liver Disease Updated for the Modern Era. Gastroenterology. 2021;161:1887–95.e4. https://doi.org/10.1053/j.gastro.2021.08.050

80 

Buchta C, Salle BD, Marrington R, Albarède S, Badrick T, Bicker W, et al. Behind the scenes of EQA–characteristics, capabilities, benefits and assets of external quality assessment (EQA): Part IV – Benefits for participant laboratories. Clin Chem Lab Med. 2025;63:879–97. https://doi.org/10.1515/cclm-2024-1292