Guiñón, Soler, Gisell Díaz, Fernández, Rico, Bedini, Mira, and Alvarez: Analytical performance assessment and improvement by means of the Failure mode and effect analysis (FMEA)

Introduction

In the field of healthcare, the clinical laboratory is one of the areas that has made many efforts to minimize errors that may affect patient safety. Errors made in laboratories are approximately 0.3%, which is much lower than the percentage of errors in other areas of medicine (1). Despite this, taking into account the high number of laboratory tests that are usually carried out, this percentage of error can total millions of erroneous results per year. Although it is unlikely, some of these errors can lead to adverse effects on patients (2).

Errors can occur at any step in the testing process. However, in the last few years it has been found that most of the laboratory errors occur in the extra-analytical phases. This is because the vast majority of the strategies adopted to minimize risks have been focused on reducing failures in the analytical phase, such as the design of an internal quality control (IQC) plan or the participation in external quality assessment (EQA) schemes (3). Despite these strategies, nowadays the estimated percentage of analytical errors is approximately 23% of the total errors produced in laboratories (4). This indicates that it is necessary to improve analytical performance, therefore efforts to mitigate these errors should not cease.

A requirement of the international standard for accreditation of clinical laboratories, the International Organization for Standardization (ISO) 15189:2012, is the incorporation of risk management principles aimed at patient safety (5). The last version of the ISO 9001:2015 standard, frequently implemented in clinical laboratories, has also incorporated risk management to improve the quality of the processes (6). There are ISO directives and Clinical and Laboratory Standards Institute (CLSI) documents that provide guidance on risk management (7-10). Although these standards do not indicate which methodology should be used, the failure mode and effect analysis (FMEA) is the most applied (2).

Failure mode and effect analysis is a risk management tool for identifying possible failures that can occur and for solving known errors, analysing the causes and effects of the failures, and eliminating or reducing the most relevant ones by proposing control measures (11). After identifying and listing all the possible failure modes, the risk must be estimated using different factors. Clinical and Laboratory Standards Institute EP23A “Laboratory Quality Control Based on Risk Management” recommends a 2-factor risk model considering only the occurrence and the severity of a failure. However, to assess the analytical performance in medical laboratories a 3-factor risk model can be more appropriate (12). This model considers also the detectability, as the probability of not detecting a failure. Then, a value is assigned to each risk factor according to a numerical scale. Higher scores are assigned to those risks with a higher occurrence, higher severity and lower probability of detection. To identify the high priorities for action, the values assigned to the occurrence, severity and detectability are multiplied, thus obtaining an index score named risk priority number (RPN). Risk mitigation actions are more urgently needed for those risks with a higher RPN. After the implementation of these actions, the RPN must be calculated again to assess their effectiveness (3, 11). Despite being the most used methodology, FMEA has shown some drawbacks such as the lack of standardization of scoring scales, subjectivity when assigning values to each risk factor or the lack of a unified criterion to assess the RPN value obtained (13).

It has been postulated that the Sigma metric, a tool that clinical laboratories have been using for some years to measure the effectiveness of its quality control procedures, could also be useful for the risk management of the analytical phase (14). One of the advantages of the Sigma metric is that it allows an objective assessment of the process performance and can also be used as a benchmark to compare the results obtained universally. The objective of the Six Sigma model is to reduce the variability of the processes, in such a way that six standard deviations can fit within the established tolerance limits of the process. Reducing variation, fewer defects (the results that fail to meet the specified tolerance limits) are generated so a performance improvement of the process is achieved. In order to use Sigma metric as a benchmarking scale, the number of defects produced must be expressed in defects per million opportunities (DPMO), and this must be converted into a Sigma value. The higher the Sigma value is, the lower the variation of the process. Six Sigma levels have been described (15). A Sigma value of 6, recognized as the world-class quality, corresponds to 3.4 DPMO. Sigma values between 5 and 6 are considered of excellent quality. Sigma levels of 4, 3 and 2 correspond to a good, moderate and poor quality, respectively. Sigma values below 2 indicate an unacceptable quality. It has been described that the number of defects in the analytical phase due to quality control failures (IQC and EQA) is around 3 to 4-Sigma (15). Outside of health care, a Sigma value of 3, which corresponds to 66,807 DPMO, has been described as the minimum acceptable quality for routine operation (14). Currently, as clinical laboratories perform a high number of tests, the same standards might be applied (16). Obtaining lower Sigma values would indicate the need to implement improvement actions.

The objective of this study was to carry out an FMEA of the analytical phase to assess the analytical performance of the measurement procedures, implement actions to mitigate the risks and evaluate their effectiveness.

Materials and methods

Materials

The study was performed in the Biomedical Diagnostic Center of the Hospital Clinic of Barcelona, Spain, in the period between 1st of January 2017 and 31st of December 2018. In the study were included all measurands whose measurement errors (ME) exceeded the allowable total error (TEa) in any of the surveys of the EQA scheme “Serum biochemistry” organized by the Spanish Society of Laboratory Medicine (SEQC). The following measurands were evaluated: alkaline phosphatase (ALP), direct bilirubin (DBIL), chloride (Cl), creatine kinase (CK), creatinine (CREA), high density lipoprotein (HDL), lactate dehydrogenase (LD), potassium (K), sodium (Na) and total protein (TP). The materials analysed were: i) quality control materials provided by the EQA scheme and ii) IQC materials: Liquid Assayed Multiqual level 2 and 3 (Bio-Rad Laboratories, Marnes-la-Coquette, France).

Methods

All tests were run on ADVIA Chemistry XPT System analyser (Siemens Healthcare GmbH, Erlangen, Germany). From each EQA survey, the ME were obtained. It was calculated (Equation (Eq.) 1) as the distance in percentage from the value reported by the laboratory to the target value (instrument group mean):

ME (%) = [(Result – Target value)]
x 100 / Target value

The selection of the allowable total error was preferably based on the biological variability (BV) (minimum, desirable or optimal values) (17). When the BV model was so demanding that the laboratory was not possible to achieve the minimum values, the criteria of the Clinical Laboratory Improvement Amendments (CLIA) was adopted (18). Table 1 shows the TEa set by the laboratory for the measurands included in this study.

Table 1

Total error quality specifications established by the laboratory

Measurand Source of Quality Specification TEa (%)
ALP Desirable BV 12.04
DBIL Optimal BV 22.30
Cl CLIA 5.00
CK Optimal BV 15.20
CREA Desirable BV 8.87
HDL Desirable BV 11.63
LD Desirable BV 11.35
K Minimum BV 8.40
Na CLIA 2.79
TP Minimum BV 5.40
BV – biological variability. CLIA – Clinical Laboratory Improvement Amendments. TEa – allowable total error. ALP – alkaline phosphatase. DBIL – direct bilirubin. Cl – chloride. CK – creatine kinase. CREA – creatinine. HDL – high density lipoprotein. LD – lactate dehydrogenase. K – potassium. Na – sodium. TP – total protein.

Failure mode and effect analysis of the years 2017 and 2018 were performed using a 3-factor model. In order to assign objective values to the occurrence, severity and detectability, they were calculated using data from the EQA and the IQC, as follows:

1) Occurrence (O) was calculated (Eq. 2) as the percentage of errors (results whose ME exceeded the TEa) with respect to the number of surveys of the EQA scheme:

O (%) = Number of errors x 100 /
Number of surveys

Those measurands that did not exceed the TEa in any of the EQA surveys were not included in the FMEA, since the occurrence value would have been zero.

2) Severity (S) was calculated (Eq. 3) as the difference expressed as a percentage between the ME and the TEa. A ME higher than TEa means a significant deviation that could lead to a false diagnosis or inadequate treatment, so the risk for the patient safety would rise as that difference increases.

S (%) = (ME – TEa) x 100 / TEa

When for a certain measurand there was more than one ME higher than the TEa, the average of ME was calculated.

3) Detectability (D) was calculated (Eq. 4) from the probability of error detection (Pde) provided by the Westgard Advisor module available in Unity Real Time v.2.0, a quality control data management software from Bio-Rad Laboratories (19). With this purpose, we obtained the Pde for both concentration levels of IQC and calculated the average. Since Pde is the probability of detecting an error, the detectability was calculated as the complementary value of the Pde (1 - Pde), thus obtaining the probability of not detecting an error. When the Pde value was 1, it was transformed into 0.999 in order to obtain data for the calculation. Results were expressed as a percentage.

D (%) = (1 – Pde) x 100

Following the methodology published by Sten Westgard, risk factors were expressed as a percentage to obtain the RPN on a scale of 1 to 1 million. Therefore, an estimation of the number of DPMO was obtained (15). Then, the DPMO was converted into a Sigma value (σFMEA) to standardize the evaluation of the results and the prioritization of actions, providing a better assessment of the risks that are acceptable than the one provided by the RPN. The DPMO value (described in the Six Sigma conversion tables) closest to the DPMO obtained provides the σFMEA (20).

The laboratory investigated the possible causes of failure for those measurands with a σFMEA lower than 3. Thus, a root cause analysis by means of the repeated “whys” query tool was performed (21). After the investigation, some actions were taken to decrease the number of DPMO produced, such as applying an equation to correct systematic error or increasing the frequency of IQC concentration level 3 testing (3 times a day).

An FMEA of the year 2018 was performed to evaluate the effectiveness of these measures and to identify new residual risks. A σFMEA between 3 and 4-Sigma or higher would indicate the success of the corrective measures.

Results

Tables 2 and 3Table 3 show the ME obtained from the EQA surveys in the years 2017 and 2018 respectively.

Table 2

Measurement errors expressed as a percentage obtained in the EQA surveys of the year 2017

ME (%)
Measurand Jan Feb Mar Apr May Jun Jul Aug Sep Oct Nov Dec
ALP 8.91 3.96 2.43 5.20 1.31 10.16 11.79 18.85 11.08 14.93 17.66 15.84
DBIL 14.60 30.47 24.02 1.44 4.49 4.63 4.02 0.39 5.81 0.85 7.29 0.23
Cl 1.08 3.39 1.21 2.17 1.54 0.26 1.96 0.70 0.63 2.10 1.75 7.81
CK 16.12 9.29 11.54 1.70 2.90 4.81 2.21 3.00 4.16 2.62 1.80 7.13
CREA 8.45 2.49 0.95 1.93 4.62 0.39 2.54 9.46 0.13 5.80 2.92 3.49
HDL 17.47 3.52 9.28 10.61 9.87 0.52 14.75 9.22 3.05 9.09 1.39 5.55
LD 4.79 6.03 5.39 6.15 7.33 3.68 5.52 4.37 5.49 3.95 5.42 12.61
K 1.86 1.98 0.73 3.23 0.74 1.19 1.15 0.21 1.01 0.21 1.56 11.62
Na 1.37 1.24 0.01 0.98 0.95 0.54 0.22 0.48 1.46 3.84 2.59 6.87
TP 0.35 2.34 0.97 1.38 0.65 7.07 0.31 4.65 1.84 2.93 1.00 1.78
ME – measurement error. ALP – alkaline phosphatase. DBIL – direct bilirubin. Cl – chloride. CK – creatine kinase. CREA – creatinine. HDL – high density lipoprotein. LD – lactate dehydrogenase. K – potassium. Na – sodium. TP – total protein.
Table 3

Measurement errors expressed as a percentage obtained in the EQA surveys of the year 2018

ME (%)
Measurand Jan Feb Mar Apr May Jun Jul Aug Sep Oct Nov Dec
ALP 6.01 10.55 13.45 5.14 5.85 4.61 8.66 8.25 5.38 9.38 21.09 4.35
DBIL 2.09 10.02 4.57 0.59 2.61 1.92 6.37 0.26 2.59 3.89 16.66 10.74
Cl 2.71 4.00 1.17 2.62 0.64 5.82 0.07 1.16 0.45 0.61 0.51 0.28
CK 3.64 1.63 3.21 0.54 2.94 2.29 8.27 7.40 1.59 4.76 2.34 1.52
CREA 0.74 1.90 6.92 3.13 2.68 1.22 4.49 7.47 1.90 3.98 11.65 5.85
HDL 0.27 2.37 0.88 0.39 3.35 0.53 26.50 0.57 0.13 12.37 0.90 2.53
LD 4.12 22.50 2.95 0.61 1.44 1.19 1.95 1.24 0.01 0.69 2.26 1.34
K 0.81 4.40 0.06 4.48 2.73 4.19 0.49 0.71 0.25 1.92 0.23 2.65
Na 0.46 3.80 1.61 1.31 0.65 2.84 2.36 0.53 0.09 1.19 0.05 1.55
TP 3.85 2.54 0.47 3.29 3.45 2.87 1.48 0.47 1.81 1.57 2.16 3.52
ME – measurement error. ALP – alkaline phosphatase. DBIL – direct bilirubin. Cl – chloride. CK – creatine kinase. CREA – creatinine. HDL – high density lipoprotein. LD – lactate dehydrogenase. K – potassium. Na – sodium. TP – total protein.

The results of the FMEA of the year 2017 are shown in Table 4. Ten measurands were evaluated. Alkaline phosphatase and Na presented a σFMEA of less than 3 and only CK was close to the world-class quality. The results of the FMEA of the year 2018 are shown in Table 5. Only six measurands were included. None of the measurands presented a σFMEA below 3. The results showed that σFMEA for ALP and Na increased.

Table 4

FMEA of the year 2017

Measurand Occurrence (%) Severity (%) Pde Detectability (%) DPMO σFMEA
ALP 33.33 39.70 0.200 80.00 105,856 2.7
DBIL 16.67 22.17 0.948 5.25 1940 4.4
Cl 8.33 56.20 0.500 50.00 23,407 3.5
CK 8.33 6.05 0.999 0.10 5 5.9
CREA 8.33 6.65 0.610 39.05 2163 4.4
HDL 16.67 38.52 0.431 56.90 36,537 3.3
LD 8.33 11.10 0.200 80.00 7397 3.9
K 8.33 38.33 0.904 9.65 3081 4.2
Na 16.67 91.94 0.274 72.65 111,346 2.7
TP 8.33 30.93 0.426 57.45 14,802 3.7
Pde – probability of error detection. DPMO – defects per million opportunities. σFMEA – Sigma value. ALP – alkaline phosphatase. DBIL – direct bilirubin. Cl – chloride. CK – creatine kinase. CREA – creatinine. HDL – high density lipoprotein. LD – lactate dehydrogenase. K – potassium. Na – sodium. TP – total protein.
Table 5

FMEA of the year 2018

Measurand Occurrence (%) Severity (%) Pde Detectability (%) DPMO σFMEA
ALP 16.67 43.44 0.569 43.15 31,247 3.4
Cl 8.33 16.40 0.625 37.50 5122 4.1
CREA 8.33 31.34 0.595 40.50 10,607 3.8
HDL 8.33 67.11 0.449 55.10 30,802 3.4
LD 8.33 98.24 0.999 0.10 82 5.3
Na 16.67 19.00 0.582 41.80 13,239 3.7
Pde – probability of error detection. DPMO – defects per million opportunities. σFMEA – Sigma value. ALP – alkaline phosphatase. Cl – chloride. CREA – creatinine. HDL – high density lipoprotein. LD – lactate dehydrogenase. Na – sodium.

Discussion

In the present study, a 3-factor risk model was used to perform the FMEA. The proposed methodology allowed us to assess the analytical performance of the laboratory methodologies, detecting poor analytical performance of ALP and Na, which led us to implement improvement actions. The actions taken in order to reduce the number of DPMO were different for the two measurands according to the results of the root-cause analysis. Therefore, for the bias on the EQA results of ALP from August until December 2017, that was due to a lot-to-lot reagent variation, the action taken by the laboratory was to apply an equation to correct the systematic deviation until the reagent lot changed. The elimination of the bias led to a higher probability of detection in 2018 (from 0.200 to 0.569) and a reduction of the occurrence of errors in the EQA surveys. In the case of Na, a poor performance of IQC concentration level 3 was observed. The action taken was to increase the frequency of IQC concentration level 3 testing. The modification of the quality control procedure in 2018 increased the probability of error detection (from 0.274 to 0.582) and the errors committed in the EQA surveys had a lower severity. Because of the actions carried out in 2017, σFMEA obtained for these measurands in the FMEA of the year 2018 were between 3 and 4-Sigma, indicating an improvement of the analytical performance.

Other authors have applied the FMEA methodology to assess the analytical performance of clinical chemistry measurands (22-25). As a conclusion, all of them stated the usefulness of the risk assessment to detect if it was necessary to adjust the quality control procedures in order to improve the analytical performance. However, due to the lack of standardization of the FMEA methodology, it is not feasible to compare the results of the present study with those obtained in their studies. They also used a modified FMEA and the Sigma metric in order to overcome the drawbacks of the FMEA. However, their methodology was neither entirely objective nor easy to standardize mainly because the value assigned to the severity was based on personal experience (22-24). Furthermore, those studies do not explain the risk mitigation actions proposed nor even verify the effectiveness of actions implemented. Only Capunzo et al., that performed an FMEA of the analytical process of three measurands (glucose, total cholesterol, and total bilirubin), listed the improvement actions performed and reassessed the risks once these actions were applied (25). As in the present study, they achieved a risk reduction (measured as RPN) for all measurands after implementing actions such as increasing the IQC frequency analysis or changing the procedure of preparation, storage and use of the calibrators.

It is important to remark that all the factors used to perform the FMEA depend on the TEa established by the laboratory. Therefore, in order to analyse correctly the results obtained, the laboratory must also assess whether the established tolerance limits are the right ones. Depending on the TEa established by the laboratory, it is possible to move from a σFMEA of less than 3 to a σFMEA of more than 3 or vice versa, as it occurs in other studies in which they have assessed the importance of selecting a proper TEa to obtain a reliable Sigma value that represents analytical performance (26). As it has been shown in other studies the current technology does not always allow to reach quality specifications based on BV, even though these specifications should be the objective (27). Therefore, to assess the results correctly, it should be considered whether the goals set by the laboratory are realistic or not. The evaluation of the analytical performance of the measurement procedures may not be adequate if the quality specifications established by the laboratory are too demanding. According to the authors, laboratory quality departments and responsible laboratory personnel should set together adequate goals.

The proposed methodology allowed us to identify those magnitudes with a less satisfactory analytical performance and, once the causes were detected and the improvement actions implemented, verify their effectiveness. Performing an FMEA using EQA and IQC data allows to standardize how the scoring scales are made. This means a reduction in the time needed by the staff to calculate the risk score and ensures the treatment of risks under the same criteria. Furthermore, this study shows that the Sigma metric can be used in the FMEA to standardize the prioritization of risks, as well as to measure the change in risk score after having made risk mitigation actions.

Notes

[1] Conflicts of interest None declared.

References

1 

Lippi G, Plebani M. A Six-Sigma approach for comparing diagnostic errors in healthcare—where does laboratory medicine stand? Ann Transl Med. 2018;6:180. https://doi.org/10.21037/atm.2018.04.02

2 

Aita A, Padoan A, Antonelli G, Sciacovelli L, Plebani M. Patient safety and risk management in medical laboratories: theory and practical application. J Lab Precis Med. 2017;2:75. https://doi.org/10.21037/jlpm.2017.08.14

3 

Jiang Y, Jiang H, Ding S, Liu Q. Application of failure mode and effects analysis in a clinical chemistry laboratory. Clin Chim Acta. 2015;448:80–5. https://doi.org/10.1016/j.cca.2015.06.016

4 

Plebani M. The journey toward quality and patient safety in laboratory medicine continues. N Am J Med Sci. 2014;6:229–30. https://doi.org/10.4103/1947-2714.132942

5 

International Organization for Standardization (ISO). ISO 15189:2012. Medical laboratories – requirements for quality and competence. Geneva: ISO, 2012.

6 

International Organization for Standardization (ISO). ISO 9001:2015. Quality management systems -requirements. Geneva: ISO, 2015.

7 

International Organization for Standardization (ISO). IEC 31010:2019. Risk management – risk assessment techniques. Geneva: ISO, 2019.

8 

International Organization for Standardization (ISO). ISO/TS 22367:2008. Medical laboratories – reduction of error through risk management and continual improvement. Geneva: ISO, 2008.

9 

Clinical and Laboratory Standards Institute (CLSI). Laboratory quality control based on risk management; Approved Guideline. CLSI document EP23-A. Wayne, PA: CLSI, 2011.

10 

Clinical and Laboratory Standards Institute (CLSI). Risk management techniques to identify and control laboratory error sources; Approved guideline, 2nd ed. CLSI document EP18-A2. Wayne, PA: CLSI, 2009.

11 

Liu HC. FMEA for Proactive Healthcare Risk Analysis: A Systematic Literature Review. In: Improved FMEA Methods for Proactive Healthcare Risk Analysis. Singapore:Springer; 2019. p.15-45. https://doi.org/https://doi.org/10.1007/978-981-13-6366-5

12 

Westgard JO. Perspectives on quality control, risk management, and analytical quality management. Clin Lab Med. 2013;33:1–14. https://doi.org/10.1016/j.cll.2012.10.003

13 

Dean Franklin B, Shebl NA, Barber N. Failure mode and effects analysis: too little for too much? BMJ Qual Saf. 2012;21:607–11. https://doi.org/10.1136/bmjqs-2011-000723

14 

Westgard S. Prioritizing risk analysis quality control plans based on Sigma metrics. Clin Lab Med. 2013;33:41–53. https://doi.org/10.1016/j.cll.2012.11.008

15 

Westgard JO, Armbruster D, Westgard SA. Risk, Error and Uncertainty: Laboratory Quality Management in the Age of Metrology. An Issue of the Clinics in Laboratory Medicine, Vol. 37-1. New York: Elsevier, 2017. https://doi.org/https://doi.org/10.1016/s0272-2712(16)30120-2

16 

Westgard S, Bayat H, Westgard JO. Analytical Sigma metrics: A review of Six Sigma implementation tools for medical laboratories. Biochem Med (Zagreb). 2018;28:020502. https://doi.org/10.11613/BM.2018.020502

17 

Ricos C, Alvarez V, Cava F, Garcia-Lario JV, Hernandez A, Jimenez CV, et al. Desirable specifications for total error, imprecision, and bias, derived from intra- and inter-individual biologic variation. The 2014 update. Available at: http://www.westgard.com/biodatabase1.htm. Accessed: January 2nd 2019.

18 

CLIA Requirements for Analytical Quality. Available at: https://www.westgard.com/clia.htm. Accessed October 2nd 2018.

19 

Westgard JO. Basic Quality Management Systems. Chapter 12. Designing Statistical QC procedures. Madison WI: Westgard QC; 2014. p. 171-88.

20 

Six Sigma Table. Available at: https://www.westgard.com/sixsigtable.htm. Accessed February 21th 2019.

21 

Clinical and Laboratory Standards Institute (CLSI). Nonconforming Event Management; Approved guideline, 2nd ed. CLSI guideline QMS11. Wayne, PA: CLSI, 2015.

22 

Xia Y, Xue H, Yan C, Li B, Zhang S, Li M, et al. Risk analysis and assessment based on Sigma metrics and intended use. Biochem Med (Zagreb). 2018;28:020707. https://doi.org/10.11613/BM.2018.020707

23 

Ceriotti F, Cappelletti P, Caputo M, Di Serio F, Messeri G, Ottomano C, et al. A risk-analysis approach to the evaluation of analytical quality. Clin Chem Lab Med. 2011;50:67–71. https://doi.org/10.1515/cclm.2011.740

24 

Westgard SA, Salcedo G. A Six Sigma risk analysis example: defining QC frequency with FMEA. July 2012. Available at: https://www.westgard.com/endoclab-risk.htm. Accessed October 17th 2018.

25 

Capunzo M, Cavallo P, Boccia G, Brunetti L, Pizzuti S. A FMEA clinical laboratory case study: how to make problems and improvements measurable. Clin Leadersh Manag Rev. 2004;18:37–41.

26 

Hens K, Berth M, Armbruster D, Westgard S. Sigma metrics used to assess analytical quality of clinical chemistry assays: importance of the allowable total error (TEa) target. Clin Chem Lab Med. 2014;52:973–80. https://doi.org/10.1515/cclm-2013-1090

27 

Molina A, Guiñon L, Perez A, Segurana A, Bedini JL, Reverter JC, et al. State of the art vs biological variability: Comparison on hematology parameters using Spanish EQAS data. Int J Lab Hematol. 2018;40:284–91. https://doi.org/10.1111/ijlh.12783