Ercan, Akbulut, Avcı, Yücel, Oğuz, Turhan, and Serdar: Determining biological variation of serum parathyroid hormone in healthy adults

Introduction

Parathyroid hormone (PTH) is a large polypeptide hormone initially synthesized as a pre-proprotein which later undergoes cleavages to generate the biologically active fragment containing 84 amino acid (1). It is stored in the parathyroid glands and secreted into circulation to regulate calcium and phosphate homeostasis through interacting with its receptors localised on the target cells in renal tubules, bone and intestine (2-4). Measurement of PTH has a major effect on the accurate diagnosis and treatment of related metabolic disorders (5, 6). Interpretation of PTH test results is always a challenge for both physicians and laboratory specialists. There are many factors affecting PTH results such as diurnal and seasonal rhythms and poor standardization of sample collection and differences in antibodies used by different methods. All these variations should be kept in mind to obtain reliable PTH results (7-10).

In the clinical laboratory, production and applications of data on biological variation (BV) data is essential in order to improve the quality of laboratory testing process (11, 12). For this purpose, the main measures derived from BV data and used in the laboratory practice are determination of analytical goals, evaluation of population based reference values for convenience and assessment of the significant difference between serial results belonging to the same individual (13-15). Components of BV are: i) within-subject biological variation (CVI) which is defined as the random fluctuations around a set point within the same individual and ii) between-subject biological variation (CVG), which is defined as the variations occurring due to different equilibrium points of different individuals. Calculation of CVI and CVG values is needed for the definition of desirable specifications (11, 12, 16).

Presentation of new data on CVI and CVG values, setting goals for accuracy, precision and total acceptable error derived from these BV estimates for an analyte will contribute to the literature and clinical laboratory practice in terms of process improvement. This is the first study in Turkey related to the biological variation of PTH. To the best of our knowledge, there are only three pre-existing reports related with BV of PTH in literature which were conducted in the United Kingdom (UK) (17-19). In previous studies, plasma was the specimen type while we used serum samples as extensively done in our country. We also carried on sampling for a longer period of time (10 weeks) than the other reports. Additionally, PTH data stratified according to gender was missing in the literature.

The aim of the present study is to determine the BV of serum PTH. We were also intended to contribute to clinical BV applications used in health care systems.

Materials and methods

Subjects

This prospective study was conducted at Ankara Numune Training and Research Hospital Biochemistry Laboratory, Ankara, Turkey. Twenty healthy volunteers with 10 male (median age 34 years, range: 25-54) and 10 female (median age 32 years, range: 23-40) subjects were included. The participants were in completely healthy status and they did not use any kind of medication or herbal supplement. The exclusion criteria were having concomitant autoimmune or autoinflammatory disease, acute or chronic infection, malignancy, systemic diseases such as diabetes mellitus and heart failure, pregnancy or being in the postpartum 6 months, kidney disease, undergone parathyroidectomy or bone disease. The participants were advised to maintain their normal lifestyles while the study was conducted. The study protocol was approved by the Ethics Committee of Ankara Numune Training and Research Hospital (Ref number: 1529/2017) and written informed consent was obtained from each participant.

Methods

Peripheral venous blood specimens were drawn according to the recommendations of the Clinical Laboratory Standards Institute (CLSI), Document GP41 (Collection of Diagnostic Venous Blood; Approved guideline, 7th ed.) in order to minimise preanalytical variability. Subjects underwent phlebotomy between 9-10 in the morning after a fasting period of 8 hours by the same phlebotomist on the same day of the week for 10 consecutive weeks between April 2017 and June 2017. The samples were collected in serum vacuum tubes containing separator gel (volume of 5 mL) BD Vacutainer SST II advance (Ref number: 8260910) (Becton, Dickinson and Company Franklin Lakes, NJ, USA) and sera were obtained after centrifugation at 1500xg for 10 minutes. Afterwards, serum samples were aliquoted into Eppendorf tubes and kept at - 80°C until analysis. Serum concentrations of calcium and phosphate were analyzed on biochemistry autoanalyzer Cobas c702 (Roche Diagnostics, Mannheim, Germany) and PTH concentrations were analyzed on Cobas e601 electrochemiluminescence immunoassay system (Roche Diagnostics, Mannheim, Germany) by 2nd generation Elecsys PTH assay.

All samples belonging to the same volunteer were assayed in the same batch for minimising inter-batch analytical variation. All samples and internal quality control materials were analyzed in duplicate. We conducted an internal quality control process using 2 levels of PreciControl Varia (Ref number: 05618860) (Roche Diagnostics, Mannheim, Germany). Internal quality control process was within acceptable ranges during the study period. The same lots of calibrators, reagents and quality control materials were used during the study and all analyses were performed by the same analyst.

Statistical analysis

Prior to analysis, outliers were identified and removed (Bartlett and Cochran tests). Subsequently, confirmation of the outliers was done by visually inspection a graph of data (scatter plot diagram). The Anderson-Darling test was used to check the normality of within- and between-subject data. Data were expressed as the mean ± standard deviation for normally distributed data. The differences between mean values in female and male groups were tested using Student’s t-tests. The values P < 0.05
were considered statistically significant. To calculate CVI, CVG and analytical variation (CVA) nested analysis of variance (nested ANOVA) was used.

Analytical variation was calculated from the duplicate results of each sample. Standard deviation (SD), CVA, CVI, CVG, individuality index (II) and reference change value (RCV) were calculated according to the formulas described by Fraser and Harris as follows (20):

SDA2 = (∑d2/2n) and CVA = (SDA/Mean) x 100
CVI = (CVTI 2 – CVA 2)1/2
CVG = (CVT2 – CVI2 – CVA2)1/2
II = CVI / CVG
RCV = 21/2 x Z x (CVA2 + CVI2)1/2,
where SDA represents the analytical standard deviation; d the difference between duplicates; n the number of paired result; CVTI the total within subject variation, and Z the probability selected for statistical significance (Z = 1.96 at 95% confidence interval, CI).

Biological variation estimates were used to calculate analytical performance specifications (APS) for analytical imprecision, for analytical bias and for total allowable error. Data analysis was performed using Analyse-it for Microsoft Excel 4.0 (Analyse-it Software Ltd., Leeds, UK) and XLSTAT® software (Addinsoft, Paris, France).

Results

Distribution of serum calcium (2.3 ± 0.1 mmol/L) and inorganic phosphate (1.2 ± 0.2 mmol/L) concentrations were given in Table 1. We included 371 measurements for all calculations from the 400 results. Total of outliers was 7.25%.

Table 1

Serum concentrations of the measured analytes

Subjects
(population based reference intervals)
All group
(N = 20)
Male
(N = 10)
Female
(N = 10)
P
PTH
(15-65 ng/L)
34.6 ± 10.9 38.2 ± 11.7 31.1 ± 8.8 < 0.001
Calcium
(2.1-2.5 mmol/L)
2.3 ± 0.1 2.4 ± 0.1 2.3 ± 0.2 0.500
Inorganic phosphate
(0.8-1.5 mmol/L)
1.2 ± 0.2 1.3 ± 0.1 1.1 ± 0.2 0.200
Data are presented as mean ± standard deviation. PTH - parathyroid hormone. P < 0.05 was considered statistically significant.

Table 2 lists CVA, CVI, CVG with the corresponding 95% coefficients of variation (CIs), II and RCV for PTH. There were no significant differences in 95% CIs for CVI and CVG estimated between females and males. Serum PTH distributions for both genders were given as box plot graph in Figure 1.

Table 2

Variance components for PTH derived from BV data

Male Female All group
CVA, % (95%CI) 3.8 (3.3-4.5)
CVI, % (95%CI) 18.5 (15.8-21.5) 24.0 (21.5-27.6) 21.1 (19.4-23.2)
CVG, %(95%CI) 26.2 (17.5-48.7) 18.6 (11.6-35.8) 24.9 (18.4-37)
II 0.7 1.2 0.8
RCV 52.3 67.3 59.4
CVA was calculated using the duplicate results of participants. CVA - analytical coefficient of variation. CI - coefficient of variation. CVI - within-subject biological variation. CVG - between-subject biological variation. II - individuality index. RCV - reference change value
Figure 1

Distribution of iPTH concentration according to gender

bm-29-3-030702-f1

The mean values of PTH results for two genders were shown in Figure 2. Three level model for APS including minimum, desirable and optimal goals for precision, bias and total error for PTH derived from the biological data for all group were shown in Table 3.

Figure 2

Mean with range (minimum-maximum) of iPTH test for individuals based on weekly samplings for 10 weeks

bm-29-3-030702-f2
Table 3

Analytical performance specifications for PTH measurement derived from our data on BV for all group

Quality level Precision Bias TEa
Optimal performance, % < 5.3 < 3.1 < 11.8
Desirable performance, % < 10.6 < 6.3 < 23.7
Minimal performance, % < 15.8 < 9.4 < 35.5
TEa - total allowable error. CVA - analytical coefficient of variation. CVI - within-subject biological variation. CVG - between-subject biological variation. Optimal performance: Imprecision CVA < 0.25 CVI, Bias < 0.125 (CVA2 + CVG2)1/2, TEa < 0.125 (CVA2 + CVG2)1/2 + 1.65 (0.25 CVI). Desirable performance: Imprecision CVA < 0.50 CVI, Bias < 0.250 (CVA2 + CVG2)1/2, TEa < 0.250 (CVA2 + CVG2)1/2 + 1.65 (0.5 CVI). Minimal performance: Imprecision CVA < 0.75 CVI, Bias < 0.375 (CVA2 + CVG2)1/2, TEa < 0.375 (CVA2 + CVG2)1/2 + 1.65 (0.75 CVI).

Discussion

The present study revealed that PTH has high within and between individual biological variations with low individuality index in healthy adults. We also detected that CVı and CVG values were similar for both gender. In the study by Ankrah-Tetteh et al. the results were similar to our findings, as CVI and CVG were 25.9 and 23.8, respectively (17). Viljoen et al. planned their study on twenty healthy volunteers and revealed that CVI and CVG for PTH were 25.3% and 43.4%, respectively (18). While the CVI result was similar to our finding, the CVG was higher than our result. Gardham et al. reported a value of 19.2% for CVI from a group including 12 healthy participants (19).

This is the first report providing BV parameters of PTH with 95% CIs separately for both genders and all subjects. The CIs of the CVI and CVG values were not reported previously, rendering a comparison of BV data difficult between studies. Lack of gender-stratification may limit the utility of BV estimates. In this study, APS were calculated using data from all participants since 95% CIs of CVI and CVG values according to genders overlapped.

For estimates of components of biological variation to be valid, the analytical component of variation has to be less than half of the intraindividual variation (21). Results of the other previous studies and our study are less than half of CVI. In order to achieve and maintain quality together with the generation of reliable test results in the clinical laboratory performance targets are required (11, 22). The desirable precision, bias and total allowable error for serum PTH determination was < 10.6%, < 6.3% and < 23.7%, respectively, according to the BV data obtained in this study.

In the clinical practice interpretation of a test result’s significance according to the reference interval may not ensure an accurate evaluation all the time. Individuality index is the preferred marker used to check the convenience of reference interval for result evaluation. When II of a test is less than 0.6, the distribution of the values of any individual is expected to be presented only in a small portion of the interval which means significant individuality. In such a situation, classical reference values will not be useful for the evaluation of results. If II > 1.4, the use of reference range is acceptable but if II < 1.0 it is not appropriate (22). According to our results mean PTH concentration in females was lower than that of males (P < 0.001). However, 95% CIs showed the insignificant difference between genders. CVI was higher among females while CVG was higher among males. Calculated II values as 0.8 for all participants; 0.7 for males and 1.2 for females. These values indicated that PTH results should not be evaluated according to reference intervals especially for men mostly depending on the relatively smaller intraindividual variation which meant that any small change between the consecutive measurements, while the results were still in the reference interval, may be reflecting a significant change in the clinical course. Viljoen et al. reported II 0.5 in their study for all genders (18). So their findings support ours that PTH results should be evaluated in terms of clinical outcomes, gender and other factors affecting biological variation. Reference change value is another related measure derived from BV data. The difference between consecutive measurements of an individual can be detected before the reference interval is exceeded via using RCV. For many analytes utility of RCV concept may be more helpful in the clinical decision-making process rather than the population-based normal interval as a reference interval (14). Calculated RCVs in our study for all individuals, females and males in 95% confidence interval were 59.4%, 67.3% and 52.3%, respectively. With a better analytical precision probability of changes being meaningful will be higher which means a more sensitive RCV.

Hormone parameters are hard to interpret because of poor standardization and harmonization among methodologies. This challenging situation may also be leading to a discrepancy between BV estimates when different generations of PTH assay and immunoassay systems are used. All of the previous studies used second generation “intact” PTH assay on various systems (17-19). In the present study, PTH concentrations were analyzed on Cobas e601 electrochemiluminescence immunoassay system (Roche Diagnostics, Mannheim, Germany) by 2nd generation Elecsys PTH assay.

In this study, serum was the preferred specimen for PTH just like the routine practice in our laboratory. Ankrah-Tetteh et al. also used serum specimens while Viljoen et al. and Gardham et al. conducted their study on plasma samples for PTH examinations. In the daily laboratory practice using serum has an advantage like making the analyses of calcium and phosphate possible from the same sample in addition to PTH, whereas plasma is supposed to have better stability for PTH (7).

According to Fraser and Harris components of variation can be assessed from a relatively small number of specimens collected from a small group of subjects over a reasonably short period of time (20). To our knowledge, only Røraas et al. have reported about the effect of subject number on the reliability of biological variation studies and calculated power of a variety of experimental designs with varying ratios between the analytical imprecision and the within-subject variation (23). Accordingly, our study’s power is 1.00. In a review by Ricos et al. the number of subjects included and samples were taken stated to be not very important, but at least 10 subjects and 5 samples were recommended (13). In the light of all these recommendations, samples collected from 20 participants throughout 10 consecutive weeks may be accepted as a quite sufficient sample size for a BV study.

One limitation of our study is the lack of knowledge about vitamin D status of the participants. The correlation between serum vitamin D and serum PTH concentrations is a well-known relationship. But we included healthy volunteers in the study so vitamin D deficiency may not be clinically overt. Finally, variation sources like sample type are needed to be considered while interpreting PTH results. According to the results of this study, we do not recommend the use of population-based reference intervals, furthermore RCV may be helpful in the evaluation of consecutive results of serum PTH measurement.

In conclusion, this study may contribute to BV data on serum PTH as it includes a sufficient number of volunteers from both genders over an acceptable period of time.

Notes

[1] Conflicts of interest None declared.

References

1 

Murray TM, Rao LG, Divieti P, Bringhurst FR. Parathyroid hormone secretion and action: evidence for discrete receptors for the carboxyl-terminal region and related biological actions of carboxyl-terminal ligands. Endocr Rev. 2005;26:78–113. https://doi.org/10.1210/er.2003-0024

2 

D’Amour P, Rakel A, Brossard JH, Rousseau L, Albert C, Cantor T. Acute regulation of circulating parathyroid hormone (PTH) molecular forms by calcium: utility of PTH fragments/PTH(1-84) ratios derived from three generations of PTH assays. J Clin Endocrinol Metab. 2006;91:283–9. https://doi.org/10.1210/jc.2005-1628

3 

Testini M, Gurrado A, Lissidini G, Piccinni G, Greco L, Basile F, et al. Parathyroid hormone-related peptide and primary hyperparathyroidism. Front Biosci (Schol Ed). 2010;2:299–312. [Scholar edition]https://doi.org/10.2741/s65

4 

D’Amour P, Brossard JH. Carboxyl-terminal parathyroid hormone fragments: role in parathyroid hormone physiopathology. Curr Opin Nephrol Hypertens. 2005;14:330–6. https://doi.org/10.1097/01.mnh.0000172718.49476.64

5 

Cusano NE, Maalouf NM, Wang PY, Zhang C, Cremers SC, Haney EM, et al. Normocalcemic hyperparathyroidism and hypoparathyroidism in two community-based nonreferral populations. J Clin Endocrinol Metab. 2013;98:2734–41. https://doi.org/10.1210/jc.2013-1300

6 

Ramasamy I. Inherited disorders of calcium homeostasis. Clin Chim Acta. 2008;394:22–41. https://doi.org/10.1016/j.cca.2008.04.011

7 

Cavalier E, Plebani M, Delanaye P, Souberbielle JC. Considerations in parathyroid hormone testing. Clin Chem Lab Med. 2015;53:1913–9. https://doi.org/10.1515/cclm-2015-0314

8 

Herfarth K, Schmidt-Gayk H, Graf S, Maier A. Circadian rhythm and pulsatility of parathyroid hormone secretion in man. Clin Endocrinol (Oxf). 1992;37:511–9. https://doi.org/10.1111/j.1365-2265.1992.tb01482.x

9 

Hanon EA, Sturgeon CM, Lamb EJ. Sampling and storage conditions influencing the measurement of parathyroid hormone in blood samples: a systematic review. Clin Chem Lab Med. 2013;51:1925–41. https://doi.org/10.1515/cclm-2013-0315

10 

Joly D, Drueke TB, Alberti C, Houillier P, Lawson-Body E, Martin KJ, et al. Variation in serum and plasma PTH levels in second-generation assays in hemodialysis patients: a cross-sectional study. AJKD. 2008;51:987–95. https://doi.org/10.1053/j.ajkd.2008.01.017

11 

Perich C, Minchinela J, Ricos C, Fernandez-Calle P, Alvarez V, Domenech MV, et al. Biological variation database: structure and criteria used for generation and update. Clin Chem Lab Med. 2015;53:299–305. https://doi.org/10.1515/cclm-2014-0739

12 

Fraser CG. Inherent biological variation and reference values. Clin Chem Lab Med. 2004;42:758–64. https://doi.org/10.1515/CCLM.2004.128

13 

Ricós C, Perich C, Minchinela J, Álvarez V, Simón M, Biosca C, et al. Application of biological variation – a review. Biochem Med (Zagreb). 2009;19:250–9. https://doi.org/10.11613/BM.2009.023

14 

Ricós C, Cava F, Garcia-Lario JV, Hernandez A, Iglesias N, Jimenez CV, et al. The reference change value: a proposal to interpret laboratory reports in serial testing based on biological variation. Scand J Clin Lab Invest. 2004;64:175–84. https://doi.org/10.1080/00365510410004885

15 

Fraser CG. Making better use of differences in serial laboratory results. Ann Clin Biochem. 2012;49:1–3. https://doi.org/10.1258/acb.2011.011203

16 

Røraas T, Stove B, Petersen PH, Sandberg S. Biological Variation: The Effect of Different Distributions on Estimated Within-Person Variation and Reference Change Values. Clin Chem. 2016;62:725–36. https://doi.org/10.1373/clinchem.2015.252296

17 

Ankrah-Tetteh T, Wijeratne S, Swaminathan R. Intraindividual variation in serum thyroid hormones, parathyroid hormone and insulin-like growth factor-1. Ann Clin Biochem. 2008;45:167–9. https://doi.org/10.1258/acb.2007.007103

18 

Viljoen A, Singh DK, Twomey PJ, Farrington K. Analytical quality goals for parathyroid hormone based on biological variation. Clin Chem Lab Med. 2008;46:1438–42. https://doi.org/10.1515/CCLM.2008.275

19 

Gardham C, Stevens PE, Delaney MP, LeRoux M, Coleman A, Lamb EJ. Variability of parathyroid hormone and other markers of bone mineral metabolism in patients receiving hemodialysis. CJASN. 2010;5:1261–7. https://doi.org/10.2215/CJN.09471209

20 

Fraser CG, Harris EK. Generation and application of data on biological variation in clinical chemistry. Crit Rev Clin Lab Sci. 1989;27:409–37. https://doi.org/10.3109/10408368909106595

21 

Ricós C, Alvarez V, Perich C, Fernandez-Calle P, Minchinela J, Cava F, et al. Rationale for using data on biological variation. Clin Chem Lab Med. 2015;53:863–70. https://doi.org/10.1515/cclm-2014-1142

22 

Fraser CG, editor. Biological variation: from principles to practice. Washington, DC: AACC Press, 2001.

23 

Røraas T, Petersen PH, Sandberg S. Confidence intervals and power calculations for within-person biological variation: effect of analytical imprecision, number of replicates, number of samples, and number of individuals. Clin Chem. 2012;58:1306–13. https://doi.org/10.1373/clinchem.2012.187781